Quick-200 multi-stations surface area analyzer
Quick series gas adsorption analyzer continues the many advantages of TOP, compact body, very small dead volume, stable vacuum, excellent cold bath liquid surface constant technology, is the best choice for research and development, quality control.
QUICK series instruments perform better in the scene of rapid detection of large quantities of samples. The instrument mainly detects the specific surface area of the sample and the pore size distribution over 2 nm. It can also detect the adsorption capacity of the sample to specific gases. The instrument is designed with a four-station common Dewar bottle to ensure no error in parallel tests and more stable test results
Basic performance parameters
Specific surface area: >0.0001m2/g
Pore size: 2nm -500nm
Aperture resolution: 0.02nm
Power: 110~220V, 50/60Hz
TECHNIQUES
- Isotherm curve
- BET (Brunauer, Emmett, and Teller) surface area
- Langmuir surface area
- t-Plot
- Alpha-s method
- The BJH (Barrett, Joyner, and Halenda) method
- Dollimore-heal adsorption and desorption curve
- Horvath-Kawazoe (H-K)
- The MP method
- DFT
- NLDFT
- Dubinin-Astakhov (D-A)
- Dubinin-Radushkevich (D-R)
- Summary reports
- Custom reports
CAPABILITIES
- Physisorption
- Gas adsorption
- Gas sorption
- Nitrogen sorption
- Surface characterization
- Pore analysis
- Pore size
- Pore size distribution
- BET surface area
SPECIFICATIONS
|
QUICK200 |
||
Analysis port(s) |
4 |
8 |
|
p0 Port(s) |
1 |
2 |
|
Number of micropore analysis stations |
NONE |
NONE |
|
Adsorbates |
N2 , Ar, Kr, H2 , O2 , CO2 , CO, NH3 , CH4 , etc. |
||
Surface Area Range |
0.0001 m2/g to unknown upper limit; Measurement accuracy (standard sample) ≤ ± 1.0 % |
||
Pore Size Range |
2nm – 500 nm(Options) Repeatability: ≤ 0.2 nm in mesopore range and ≤ 0.02 nm in micropore range |
||
Minimum Pore Volume |
0.0001 cm3/g |
||
Pressure Sensors (Individual sensor systems for each analysis port can be selected) |
Four high precision 1000torr sensors for analysis ports. extra 1000 Torr for p0 determination |
||
Pressure Sensor Accuracy |
± 0.15 % (Full Scale) |
||
Degassing Stations |
4 insitu |
Standard independent degassing station |
|
Degassing Temperature |
Room temperature to 400 °C (optional 500 °C), accuracy 1 °C |
||
Vacuum System |
Standard mechanical oil pump (ultimate vacuum 103 Pa) |
||
Independent vacuum system for each analysis port |
√ (Options) |
NONE |
|
Dimensions |
|
|
|
Temperature Requirements of Environment |
15 – 40 °C |
||
Humidity Requirements of Environment |
10 % – 90 % |
||
Power Requirements |
AC 220 V ± 20 V, 50/60 Hz, maximum power 300 W, current 5 A |
||
Steam absorption |
× |
× |
|
Independent Intake Gas System for Each Analysis port |
√ (Options) |
NONE |